SKC

Leland Legacy Sample Pump Kit

Not yet rated

High flows from 5 to 15 L/min

Achieve flow rates that you couldn’t reach before when using a personal sample pump. The Leland Legacy provides the high flows and long run times of a vacuum-style pump in a compact, portable, and battery-operated sampler. 

Long runs for 24 hours on one battery charge

Rechargeable lithium-ion (Li-Ion) battery pack provides 24-hour run times with impactors and other sampling devices with low back pressures. * Leland Legacy is ideal for 24-hour indoor air studies and unattended ambient air sampling.

* Leland Legacy is not recommended for high back pressure applications such as asbestosclearance sampling.

Typical Run Time:

24+ hours, based on new pump with fully charged battery:

  • Sioutas Impactor at 9 L/min (approximately 13 inches water back pressure)
  • Personal Environmental Monitor (PEM) with 37-mm, 2.0-µm PTFE filter Cat. No. 225-1709‡ at 10 L/min
  • Low-volume PUF Tube at 5 L/min
  • IMPACT Sampler (PM2.5 or PM10) at 10 L/min
  • For extended run times, the pump may be operated while attached to the approved charger?
  • Low Levels
  • Longer runs at higher flows provide more sample for the enhanced sensitivity required for measuring low concentrations, particularly in indoor air environments

Low Noise

With the noise-reducing case accessory, the Leland Legacy can operate at 10 L/min and 12 inches water back pressure with a noise level of only 52 dBA.**

** Measured 3 feet (1 meter) from pump operating at 10 L/min without media

Advanced flow control 

The patented (U.S. Patent Nos. 5,892,160) internal flow sensor measures flow directly and acts as a secondary standard to constantly maintain the set flow. Set flow is achieved immediately at start-up and flow calibration is automatically maintained by built-in sensors that compensate for differences in temperature and atmospheric pressure during sampling.

Flexible programming

Manual three-button operation - No tools needed. Simply use the Leland Legacy built-in keypad to set, calibrate, and sample! Real-time sampling parameters are available at the touch of a button.

PC programmability - The ultimate in pump scheduling and record keeping. Use Leland Legacy with your PC and DataTrac for Leland Legacy Software to create complete running sequences, download sampling and calibration data, and generate sampling reports for ISO 9000, quality programs, and occupational health and safety management systems. DataTrac Software for Leland Legacy is included in the 5-pack kits and sold separately.

Applications

Use in non-explosive locations only.

  • 24-hour exposure monitoring
  • Indoor air studies
  • Particulate sampling:
    • Pesticides — EPA TO-10A and ASTM 4861
    • Bioaerosols (spore traps/cassettes)
    • Cotton dust
    • Radon decay products (ASTM D6327-98)
    • Isokinetic sampling
    • Nanoparticles with 37-mm MCE filter at 7 L/min
    • Respirable dust with 8 L/min PPI sampler
  • Environmental sampling
  • Sampling with cascade impactors
  • Spore trap sampling (VersaTrap) 
  • PM2.5 with PEM at 10 L/min (EPA IP-10A)
Flow Range5 to 15 L/min
Flow Control SystemClosed loop with patented internal flow sensor U.S. Patent No. 5,892,160
Compensation Range15 L/min at 5 inches water back pressure
10 L/min at 12 inches water back pressure
 5 L/min at 20 inches water back pressure
AccuracyFlow rate: ± 5% of set-point after calibration to desired flow
Timing: 1 min/mo at 25 C
Atmospheric Pressure: ± 0.3 in Hg
Flow FaultIf the pump is unable to compensate for > 15 seconds due to excessive back pressure, a flow fault icon displays and flashes, the pump enters Hold mode, and the pump retains historical data. Auto restart is attempted every 20 seconds up to 10 times. Adjustable with DataTrac for Leland Legacy software
Flow ControlAn internal isothermal flow sensor measures flow directly and continuously. Sensor readings are used in a flow monitoring algorithm to maintain calibrated volumetric flow. In addition, built-in atmospheric temperature and pressure sensors provide readings to correct volumetric flow for these parameters when they vary from point of calibration.
TubingRequires 3/8-inch ID tubing
DisplayLCD, displays pump serial number, pump software revision level, flow rate, volume, temperature, atmospheric pressure, time of day, run time, and pump status, i.e., hold and run as well as Setup information
Volume DisplayContinually updated, based on corrected flow rate multiplied by sampling time. When volume exceeds 99,999 liters, the pump will continue to run normally but an O_FLO error will appear on the LCD.
Time DisplayTime of day in hours and minutes (12 or 24-hour clock) with AM and PM indicators
Timer Display Range1 to 99999 minutes (69 days). If the run time exceeds 69 days, the timer display rolls over.
Operating Temperature Range32 to 113 F (0 to 45 C)
Protect sample pump from weather when in use outdoors.
AltitudeThe pump can apply correction to volumetric flow during sampling for weather-related or altitude variations from the atmospheric pressure established at calibration up to at least 7500 feet above and 5000 feet below sea level.
Typical Run Time24+ hours:
  • Sioutas Impactor at 9 L/min (approximately 13 inches water back pressure)
  • Personal Environmental Monitor (PEM) with 37-mm, 2.0-µm PTFE filter Cat. No. 225-1709‡ at 10 L/min
  • Low-volume PUF Tube at 5 L/min
  • DPS Sampler (PM2.5 or PM10) at 10 L/min
  • For extended run times, the pump may be operated while attached to the approved charger
  • 8 L/min Respirable PPI
Results obtained using a new pump and new fully charged battery. Pump and battery performance may vary.

Back pressure on PTFE filters can vary within the same lot. Maximum operating temperature is 464 F (240 C) based on support ring.
Noise Level62.5 dBA — pump without case
52 dBA — pump housed in noise-reducing case (optional accessory Cat. No. 224-89) measured 3 ft (1 m) from pump operating at 10 L/min without media
User-adjustable ValuesSample run time, calibration, clock display, flow rate, time of day, delayed start, and temperature and atmospheric pressure display
Recorded ValuesStart date and time, stop date and time, total sample time, flow rate, sample volume, temperature, atmospheric pressure, and pump mode transitions
Adjustable LoggingIntervalRecords pump history from 3 seconds (15.4 min of data) up to 8 hours (over 102 days of data) depending on setting. Option available when using DataTrac Software
Power SupplyBattery: removable, rechargeable lithium-ion (Li-Ion) battery pack 7.4 V x 12-Ah capacity, 88.8 Wh. See Li-Ion Battery Shipment below.
Charger/AC adapter: input voltage 100-240 V AC
Battery Recharge Time (with SKC-approved charger)15 hrs
Temperature:Charging32 to 113 F (0 to 45 C)
Temperature:Storage-4 to 95 F (-20 to 35 C)
Size8 x 3.9 x 2.6 in (20 x 10 x 7 cm)
Weight36 oz (1 kg)
CaseThermoplastic with soft rubber overmolding
RFI/EMI ShieldingCE marked

Please Enter Your Order Info

Filter by:

  • Clear Filters
Product Detail
Thomas No.
1185B66
Mfr. No.
100-3002K5
Description
Leland Legacy 5-Pack Sample Pump Kit with Multi-Charger