•Molded-acrylic column (25/8 in. O.D. x 113/8 in.) filled with 11/4 pounds of 8-mesh indicating Drierite.
•Water-vapor capacity: 50 g.
•Safe for working pressures up to 90 psig.
•Flow rate should be 200 L/hr or 0.1 scfm for maximum efficiency.
•Regenerate by heating granules on a tray for 1 hr at 200 C.
•Hose barbs will fit 1/4 to 3/8 in. tubing.
Legal Information: Drierite is a trademark of W.A. Hammond Drierite Co., LTD